RAD 115 FINAL test 1-2

About this set

Created by:

Kcraft2384  on April 9, 2012

Subjects:

Radiology

Classes:

radiology

Log in to favorite or report as inappropriate.
Pop out
No Messages

You must log in to discuss this set.

RAD 115 FINAL test 1-2

A standard pattern made to measure the resolution of digital display systems is called __________ pattern
SMPTE
1/56
Preview our new flashcards mode!

Study:

Cards

Speller

Learn

Test

Scatter

Games:

Scatter

Space Race

Tools:

Export

Copy

Combine

Embed

Order by

Terms

Definitions

A standard pattern made to measure the resolution of digital display systems is called __________ pattern SMPTE
Medical images are transferred using the _______ standard DICOM
The consistent appearance of transferred images on any compatible digital display device is achieved through ______ _______ perceptual linearization
The quality control standard for FPDM was released by _____ in 1998 VESA
The task group report 18 is a procedure name for _______________ testing digital diagnostic device
A photometer is a _____ ______ luminance meter
A telescope photometer is used as a distance of _____ 1 m
Digital display relfection and ambient light conditions are assessed with ______ ______ illuminance meter
Luminance response can be tested with a ____ and _____ photometer near and telecopic
When evaluating the test pattern for geometric distortion all lines should appear _________ straight
Specular reflection can be tested with the ______, ______. ______ TG AD pattern, ambient light turned on, monitor turned off
Spatial resolution of digital display evaluated with ______ TG 18 CX
Display noise measured using the _____ test pattern TG 18 AFC
Daily evaluation of the monitors should be done with the _____ test tool TG 18 QC
A continued quality control program for digital display device should include_____, ____, ____, ____ acceptance testing, routine QC, periodic review, annual post review
what are 3 classes of digitial image artifacts image receptor, software, objects
ghost image imcomplete erasure
preprocessing corrects for what dead pixels
Flatfielding corrects for irregular testing patterns
The size and number of digital images compressed to permit _____, _____ archiving and transferring
Object artifacts in digital images are due from xray beam collimation, histogram selection, patient positioning
Backscatter radiation artifact can occur due to highly sensitive receptors
A histogram is frequency of ccurence vs. digital value
Look up table histogram is done by the radiologic tech before the exam
Collimating all 4 sides prevents histogram analysis errors
The 18 gauge hollow______needle with a stylet, developed in 1953, is used to puncture the femoral artery. Seldinger
The most common risk for the angiography patient is ___________ bleeding at the puncture site
The xray tube used for interventional radiology procedure has a _______ small target angle
Interventional radiology requres xray equipment with a _______ high power rating
The focal spot used for magnification of small vessels cannot be larger than ______mm 0.3
The size and construction of the _____determines the anode heat capacity anode disk
The power rating for an interventional radiology tube should be at least _____kW 80
A fenestrated catheter is ______ one with side holes at tip
When applied to contrast media, what does osmolarity refer to? concentration of ions
The ____artery is the one most often accessed for arteriograms femoral
A patient must have ______ prior to having an angiograph or interventional procedure a history and physical exam, order IV hydration, a diet of clear liquids, EKG, BP, pulse
The use of _____reduces the rick of drug reaction during angiographic procedures nonionic contrast
_______is an example of an IR procedure Angioplasty
Which physicist first demonstrated the principle of CT in 1970? Godfrey Hounsfield
Each CT projecton records variations in ____ and ____ mass densities and effective atomic number
The cross section of anatomy is translated in _____during CT reconstruction matrix values
The first generation of CT used _____ single detector
The reduction in _______ was principle advantage of second generation CT scanner scanning time
The third generation of CT scanners allowed for further reduction in ____ and _____ scanning time and scattered radiation
The new developement in 4th generation CT scanners is the ____ stationary detector array
Many Ct xray tubes have the capability for millions of _____ with each exam heat units
The predetector collimator reduces scatter radiation from reaching the ______ detector array
The patient dose in CT is determined by the _______ collimator prepatient
CT scanners operate on _____ high frequency power
A software program designed to do many computations simultaneously refers to _____ algorithm
Each pixel of info in CT image contains numerical information in _____unit hounsfield
The pixel size is reduces when the _____size is increased and the _____ size is fixed matric, fov
____ technology makes multislice spiral CT possible slip ring
The CT scanner must be calibrated so that _____is at a CT number zero water
Image noise in CT scanning depends on _____ pixel size, kVp, filtration, detector efficiency, slice thickness, pt dose
Pixel Size equation FOV/matrix (convert cm to mm) 1cm = 10mm

First Time Here?

Welcome to Quizlet, a fun, free place to study. Try these flashcards, find others to study, or make your own.

Set Champions

There are no high scores or champions for this set yet. You can sign up or log in to be the first!

Completed “Learn” mode

Kcraft2384