e stim/step 6: electrodes
Order by
12 terms
Terms | Definitions |
|---|---|
size | -small areas should have small electrodes and viseversa |
small electrodes | -increase current density and skin impedance-may produce pain at lower intensity -enhance depth penetration |
large electrodes | -larger areas-more comfortable -greater dispersion -more superficial |
electrode types | -carbon rubber-self adhesive |
carbon rubber | -can use gel-WET SPONGE -wet 4x4 -dry sponge -greatly decreased conductivity of the electrode |
self adhesive | -good for long term pt returning often-ENSURE GOOD CONTACT AT ALL SURFACES AND ALL ELECTRODES |
spacing | closer together: superficial flow of current (less impedance with short conductor)-far apart: deeper current flow -FOR DEEPEST PENETRATION PLACE ELECTRODES ON EITHER SIDE OF JOINT -ALLOW 1/2 TO 1 ELECTRODE SPACE BETWEEN ELECTRODES |
position | ALWAYS PLACE ELECTRODE IN LINE WITH DIRECTION OF MUSCLE BELLY |
monopolar set up | -small and big electrode-one over targeted and other over non targeted -focus treatment w/ smaller |
bipolar | -SAME SIZE-both over targeted area |
good conductors | -nerve and muscle-75% water - directly proportionalize |
bad conductors | -fat 15% water-skin 16% -bone 5% |