e stim/step 6: electrodes

size
-small areas should have small electrodes and viseversa
1/12
Preview our new flashcards mode!

Study:

Cards

Speller

Learn

Test

Scatter

Games:

Scatter

Space Race

Tools:

Export

Copy

Combine

Embed

Order by

Terms

Definitions

size -small areas should have small electrodes and viseversa
small electrodes -increase current density and skin impedance
-may produce pain at lower intensity
-enhance depth penetration
large electrodes -larger areas
-more comfortable
-greater dispersion
-more superficial
electrode types -carbon rubber
-self adhesive
carbon rubber -can use gel
-WET SPONGE
-wet 4x4
-dry sponge
-greatly decreased conductivity of the electrode
self adhesive -good for long term pt returning often
-ENSURE GOOD CONTACT AT ALL SURFACES AND ALL ELECTRODES
spacing closer together: superficial flow of current (less impedance with short conductor)
-far apart: deeper current flow
-FOR DEEPEST PENETRATION PLACE ELECTRODES ON EITHER SIDE OF JOINT
-ALLOW 1/2 TO 1 ELECTRODE SPACE BETWEEN ELECTRODES
position ALWAYS PLACE ELECTRODE IN LINE WITH DIRECTION OF MUSCLE BELLY
monopolar set up -small and big electrode
-one over targeted and other over non targeted
-focus treatment w/ smaller
bipolar -SAME SIZE
-both over targeted area
good conductors -nerve and muscle
-75% water
- directly proportionalize
bad conductors -fat 15% water
-skin 16%
-bone 5%