Radiation Physics -- Radiation Oncology
About this set
Created by:
minidoc2 on November 14, 2011
Subjects:
Classes:
Radiation Physics - Wake Forest, Class of 2012
Log in to favorite or report as inappropriate.
Order by
480 terms
Terms | Definitions |
|---|---|
Electron output depends on... | Jaw Settings |
Embryo-Fetus Exposures | Total Dose Equivalent = 5mSV = 0.5remMonthly Dose Equivalent = 0.5mSV = .05rem Per gestational period |
KERMA | Kinetic Energy Released in MatterKERMA = E transferred from beam to medium / mass |
LET | Linear Energy TransferHighest for charged particles |
1SV = ? rem | 1Sv = 100 rem |
I-131 Therapy Dose | > 30 mCi |
Proton SRS | High E (150-250 MeV)Dose modified by spreading range and modifying shape Best Dose Distribution |
Dose @ 10cm from field edge | 1% of prescribed dose |
Public Exposures- Continuous/Frequent Exposures - Infrequent Exposures - Students | Continuous/Frequent exposure = 1mSV = 0.1 remInfrequent Exposures = 5mSv = 0.5 rem Student = 1mSv = .1 rem |
Occupational Exposures- Dose Equivalent Limit - Lens of Eye - All other organs Cumulative Exposure | - Dose Equivalent Limit = 50mSv = 5rem- Lens of Eye = 150 mSv = 15 rem - All other organs = 500mSv = 50 rem Cumulative Exposure = 10mSv x age = 1rem x age |
Exposure Rate ConstantsCo-60 Rn-222 Ra-226 Ir-192 Cs-137 Au-198 Pd-103 I-125 | Co-60 = 13.07Rn-222 = 10.15 Ra-226 = 8.25 Ir-192 = 4.69 Cs-137 = 3.26 Au-198 = 2.38 Pd-103 = 1.48 I-125 = 1.46 |
Inverse square law | ... |
Treated Volume | Volume enclosed by an isodose line that adequately covers the PTV |
Characteristic x-rays result from... | Vacancies in an electron shell |
Vascular Brachy Rx Point... | 2mm |
Effective energy of an x-ray beam | About 1/2 to 1/3 of the kVp |
Temperature/Pressure Correction | ... |
Less than 15 cm separation -- what energies? | Co-60, 4-6 MV |
Backscatter Factor | Dm Phantom / Dm AirIncreases with E then decreases Increases with field size |
Roentgen applies to energies up to... | 3 MeVIonization in air only |
Roentgen to C/kg | ... |
Retina TD5/5 (whole | 4500cGy |
Skin dose increases when ... | - SSD is decreased- Bolus is used - Field size is increased - Beam is at an oblique incidence |
Cerrobend | - Melts at 70 degrees C (158 degrees F)- Harder than lead - Bismuth (most), lead, tin cadium (least) |
Increase in dose to tissues beyond healthy lung (orthovoltage) | + 10% per cm lung |
Exposure | Roentgen |
LDR Dose Rate | 40 - 200 rad/hr |
Activity Equation | ... |
IORT is collimated with... | - Aluminum collimators- Lucite cones |
Sievert is a unit of ... | Equivalent Dose |
Advantages of EPID images | - Improved latitude- Improved constrast - Real-time image display - Digital image storage |
MRI accuracy | 2 mm |
Highly Sensitive Tumors | - Oat Cell (Small, round blue cells)- Lymphoma |
rem conversion to Sv | 100rem = 1 Sv1rem = 10mSv 1mrem = 10 microSV 1rem = 1cGy = 10mSv |
Depth dose is a Function of... | 1. Energy increase = DD increase2. Field size increase = DD increase 3. Depth increase = DD DECREASE 4. SSD increase = DD increase |
Exposure Rate Constant Co-60 | ... |
Nuclear radiation with shortest range in tissue | Alpha |
Time to decay to 1% radioactivity | 6.64 half-lives |
Lateral scattered 23MV electrons travel | Much more than 6MV secondary electrons |
Optic Chiasm TD 5/5 (whole) | 50 Gy; RTOG protocols go up to 54-55Gy |
Brachial Plexus TD 5/5 (whole) | 60 Gy |
Secondary emissions from photoelectric | - Electrons- Auger Electrons - Characteristic X-rays - Target is BOUND electron |
Use Factor (U) | Fraction of operating time during which radiation is directed towards a barrier |
Workload (W) | - Weekly dose delivered @ 1cm from source- rad/week @ 1m - (# patients/week)(dose delivered @ 1m) |
Linear Attenuation Coefficient (u) | Fraction attenuated per unit lengthu = .693 / HVL |
Exposure Rate Constant (Pd-103) | 1.48 Rcm^2 / mCi-h |
Radiation Exposure Limit (Whole Body XRT Worker) | 5rem/yr = 50mSV/yr |
OAR & Depth | Decreases with depth beyond 10cm |
Heel Effect | Intensity of diagnostic beam is less on the anode side than the cathode side. |
Choice of photon beam energy in a treatment plan is governed by... | - Depth to isocenter- Proximity of PTV to lung - Depth of PTV below surface - Neutron leakage outside the beam |
Wedge Factors | - Increase with Field Size & Depth (scatter)- Measured @ Dmax or 10cm |
Source Leak Testing | - Tested every 6 months- 0.005 uCi or more is considered leakage |
Advantages of Radiochromic Film | - Tissue Equivalent- Does not require processing - Insensitive to visible light - Response (optical density vs. absorbed dose) is more linear |
Pt A dose in GYN LDR | 50-60 cGy/hr |
Perinatal death (at or around time of birth) most likely occurs as a result of irradiation during... | Major organogenesis (21-40 days) |
LINAC SRS | - Collimation 5mm - 40mm- Multiple arcs - More than 1 isocenter - < 10 MV photons < 1mm accuracy required |
Exposure Rate Constant Ra-226 | 8.25 Rcm^2 mg-h** ONLY one to be mg-hr -- all the rest are mCi-h |
Exposure Rate Constant Cs-137 | 3.26 Rcm^2 / mCi-h |
Dose Rate Pd-103 | 20-30 cGy/hr |
Factors that influence relative dose in a medium from a brachy source... | 1. Distance from source2. Self-attenuation 3. Medium attenuation 4. Scatter build up |
1Ci = ? Bq | 3.7 x 10^10 dps (Bq) = 1 Ci |
Dimension of an atom | 1 Angstrom = 10^-10 meters |
Film | - Overresponds at low E- Great spatial resolution - H & D curves exposure/dose vs. optical density - opacity 10/1 |
As the target angle decreases, the focal spot ... | Decreases |
Coefficient for Equivalent Thickness Bone | 1.65 cm of water = 1 cm of bone |
Hot Spot for 4MV photons | 20% |
mA | - Tube current- No effect on energy or depth dose - Only affects dose rate |
Brain TD 5/5 (whole) | 45 Gy |
Non-Stochastic Effects (Deterministic) | - Has a threshold- All-or-nothing |
Alpha Decay | ... |
Increase in dose to tissue beyond health lung with 20MV x-rays | + 1% per cm of lung |
Alpha Particles | - "+2" charge- short range in air - interact quickly |
Electron Output and Inverse Square | Electron output cannot be corrected by the inverse square |
Decay Constant | 0.693 / half-life |
Gamma Knife Accuracy | 0.3mm |
B+ decay | ... |
Total Scatter Factor Scp | Measured at Dmax |
Transport Index | Indicates maxiumum dose-rate at 1.0 meter |
F = ? C | F = 9/5 C + 32 |
Planning Organ at Risk Volume | Organs at risk with an additional margin to compensate for movements. |
Energy of coherent scatter photon at 90' | .511 MeV |
Cesium-132 source decays.... | 2.3% per year (half-life = 37 years) |
Electrons lose most of their energy in soft tissue by... | Ionization and Excitation |
Transmission through intraleaf MLC | 4% |
Given dose SAD Setups | ... |
Decay Constant | ... |
Gamma Knife QA | - Interlocks daily-Output monthly -Timer monthly - Leak test semi-annually - Dose profiles annually |
Standard Deviation | ... |
Proportional Counters | - Distinguish alphas & betas- Used for neutron events (neutron shielding survey) - Count individual events |
Dose outside field is caused by... | - Patient scatter- Head scatter - Leakage |
SI unity of Activity | Becquerel |
Exposure Rate Constant I-125 | 1.46 Rcm^2/mCi-h |
Wedge Angle | WA = 90' - (hinge angle/2) |
Minimum Target Dose | Lowest absorbed dose in target area |
Max Radiation Exposure to a lends for an XRT worker | 15 Rem/yr = 150mSv/yr |
Dose outside field increases with... | - Wedges- Increased field size |
F factor | F Factor converts exposure to doseF (air) = 0.876 cGy/R |
Half Life of Cs-137 | 30 yrs |
Binding energy increases as ... | the atom decays to ground state - more stable |
Reactivity of an atom depends on... | number of electrons in the outer shell |
Public Continuous Exposure | 1mSv |
Indirectly ionizing particles... | neutrons |
Neutrons produced by LINACs | > 10 mV |
B- Decay | ... |
Mean energy of electron beam | 2.33 R50 (depth at which dose falls to 50%) |
Transmission through cerrobend | 3.5 - 5% |
Photodisintegration | - Photon in, neutron gets knocked out.- Takes ~ 7 MeV to knock out a nucleon & create a photodisintegration neutron - 10 MeV start shielding - Comes from head of machine or in patient - Neutrons become a problem at 15-20 MeV |
Irradiated Volume | Volume of tissue receiving a significant dose (50% or more of the prescribed dose) |
Termination of Spinal Cord | L2 |
Skin dose is dose below skin at | 0.5mm |
X-ray quantity is proportional to... | maS |
Effective Half-Life | ... |
Photon Energy of Ra-226 | 0.83 MeV |
HVL (mm Lead) for I-125 | 0.025mm |
Range of electrons in lead | 2 MeV/mm |
Internal Target Volume (ITV) | Internal margin added to CTV to compensate for internal motion. |
Highly Sensitive Cells | - Sperm- Erythrocytes - Crypt cells - Bone marrow |
Stochastic effects | - Random events (e.g. carcinogenesis)- Increase dose, increase probability |
Average Life | T avg - 1.44 x half-life |
SI unit of radiation exposure | Coulomb/kilogram |
The photoelectric effect is proportional to... | Z cubed + 1/E-cubed |
Gap Formula | ... |
Gross Tumor Volume (GTV) | - Gross demonstrable extent and location of tumor.- Primary tumor, metastatic lymphadenopathy, or other metastases. |
Exposure Rate Equation | Exposure Rate = Activity x Exposure Rate Constant x 1/d^2 |
Stereotactic Linac Accuracy | 1.0 cm |
Memory used to manipulate CT images | RAM |
1 byte (B) equals | 8 bits |
Rest mass of 1 amu | 931 MeV |
Geiger-Mueller Detectors | - Detect presence of radiation- Have a large signal size - Have a long dead time |
Rectum TD 5/5 (whole) | 60 Gy |
Beta particle penetration | - 0.5 cm/MeV in tissue- 800 times as far in air |
Maynord's F Factor | ... |
What is the most significant influence of dose near a brachy source? | Inverse Square |
Ra-226 to Cs-137 Conversion | 10mg Ra = 25mCi Cs-137 |
Sternal Angle/Carina | T4 |
Photon Energy Rn - 222 | 0.83 MeV |
Linac Timer Error | 0.10 Mu |
1.0 cm Pb = ? Cerrobend | 1.2 cm Cerrobend |
I-131 Radiation Type & Half Life | Beta DecayHalf-life 8.05 days |
Prostate Location | Anterior - Pubic SymphysisLateral - Greater Trochanter |
Directly Ionizing Radiation | - Protons- Alpha Particles - Beta Particles - Positrons - Photons |
When a sector is skipped in an arc technique | High dose area is shifted away from the skipped sector |
Methods of Radioisotope Production | - Separation from spent reactor fuel rods (S-137 & Sr-90)- Bombarding with neutrons in a reactor Ir-192 & Co-60 - Bombarding with protons in a cyclotron-positron emitters - Elution of metastable daughter from parent T99(m) |
Gamma Decay | Isomeric Transition (Co-60 -> Tc 99m)1. n/p same 2. n,z,a same 3. gamma ray emitted from NUCLEUS |
Photon mass attenuation coefficient varies with... | - Photon energy- Linear attenuation coefficient/density - Measured in cm^2/g |
Absorbed dose is measured in... | Rad & Gray |
Dmax 6 MeV electron | 1.2 cm |
Co-60 Leakage | - 10mR/hr MAX- 2mR/hr AVG |
Dimension of Nucleus | 10^-14m (10 to the minus 14th power) |
Photoelectric Effect | Dependent on Z-cubed up to 50 keV |
Contralateral Breast Limit | 200 cGy |
Absportion | - Process whereby energy is transferred from a photon beam to electrons in a medium |
Radiation ExposureNon-Occupational/Public Total Body | - 0.1 rem/yr- 1 mSv/yr |
Photon Energy Ir-192 | 0.38 MeV = 380 keV |
Backscatter Factor | TAR @ Dmax |
Half Life Ir-192 | 73.8 Days |
Kidney (TD 5/5 to whole kidney) | 23 Gy |
TAR, TMR, TPR depends on... | 1. Energy -- as energy increases TAR increases2. Field Size -- as field size increases TAR increases 3. Depth -- as depth increases TAR decreases |
Photon Energy Sr-90/Y-90 | 2.2 MeV Beta |
Clinical Target Volume (CTV) | - Demonstrated tumor & any other tissue with presumed tumor- Represents true extent & location of tumor |
Flattening Filter | - Flattens at d = 10 cm- Steeper for 18x than 6x - Hardens beam on CAX |
Attenuation Equation | ... |
Electron Capture | ... |
Exposure is measured in ... | Roentgen |
Average Annual Exposure Radiation Worker Total Body | 1 rem/yr10 mSv/yr |
Energies less than 10 MV, PDD @ d=10 cm decreases by... | 3-4% per cm |
X-Ray quality improves with... | Filtration |
X-Ray Contamination of Electron Beams | - Increases with energy- Caused by Bremstrahlung |
Prostate Seed RX I-125 | - Primary = 144 - 160 Gy- Boost - 120 Gy |
Lens (TD 5/5 - whole lens) | 1000 cGy |
Half-Life of I-125 | 59.4 days |
Femoral Head TD 5/5 (whole) | 5200 cGy |
Photon beam calibration is done in... | Water |
CT # of bone | 1000 cGy |
Photoelectric Effect | ... |
F-Factor | - Roentgen to rad conversion factor- Greater for high - Z materials - Greater for low energy - 0.876 for Co-60 photons in air |
Half-Life of I-131 | - 8 days- Thyroid Ablation |
A 2 bit binary number gives | four bits of information |
Standard thickness of Cerrobend in therapy | 7.5 cm |
Consequences of larger pixel format | - More memory for storage- More processing time - More time to transmit images |
Dmax 4 MV X-rays | 1.0 cm |
Compton Effect | ... |
Electron Interactions | - Ionization greater with low Z materials- Excitation - ??? - Bremstrahlung - high E & high Z |
A Beam Spoiler is used to | Increase dose in buildup region while maintaining some skin sparing |
Maximum Target Dose | - Hightest does within target- Must be greater than 2 cubic centimeters |
Electron Scatter | - Multiple Coulomb scattering- Low energy e- and high Z materials create largest amount of scatter |
Small Intestine TD 5/5 (whole) | 4000 cGy |
Increase in Dose to Tissues Beyond Health Lung (4MV x-rays) | +3% per cm lung |
Lateral scatter equilibrium in an electron beam exists when the field size is... | of the order of electron energy |
Electromagnetic Radiation With Highest Frequency | Gamma |
Total Dose Equivalent (Embryo/Fetus) | 5 mSv |
Energy of Coherent Scatter at 180' | 0.25 MeV |
Public Infrequent Exposure | 5 mSv |
HVL (mm Lead) - Pd-103 | 0.008 mm |
Isomer | - Metastable Atom- Identical atoms @ different energy levels - Photon emitted from nucleus |
Orthovoltage | - 150-500 kVp- HVL 1-4 mmCu Dmax = surface |
Gantry/Collimator | +/- 1' |
Hot Spots | - Area outside target that receives dose higher than RX dose- Must be greater than 2cm-squared |
Characteristic X-Ray Energy | Defined by energy differences of electron shells |
Umbilicus | L3-L4 = Aortic Bifurcation |
Activity | Becquerel and Curie |
Absorbed Dose | Gray + Rad |
1 Kilobyte | 2^10 bytes ; 1024 bytes |
1 Megabyte | 2^20 bytes |
OAR 6x vs. 18x | Greater for 6x |
HVL (mm Lead) Cs-137 | 5.5mm |
Pair Production | ... |
Effective Z of air in photoelectric region | 7.6 |
Surface Dose for 15 keV | 90% |
Rest Mass of Electron | 0.511MeV |
Resistant Tumors | - Melanoma- Glioblastoma |
Neutrons have a high quality factor because... | They are indirectly ionizing and transfer their energy to protons, which have a large mass and are densely ionizing |
Fetal Dose Limit | - 0.5 rem/yr = 0.05 rem/mo- 5 mSv/yr = 0.5 mSv/mo |
HVL (mm Lead) Ra-226 | 12 mm |
Dose @ 1 m from field edge | 0.1% of prescribed dose |
Dose @ 30 cm from field edge | 0.2 % of prescribed dose |
Exposure Rate Constant Rn-222 | 10.15 Rcm^2 / mCi-h |
Binding energy increases with... | Z and proximity to nucleus |
Dynamic Wedge Factors | Are a function of both the start and end jaw positions |
Surface Dose (18E & 22E) | 93% |
Electron backscatter from internal shielding | - Increases with Z- Decreases with increase in energy - is absorbed in a low Z material lining |
Suprasternal Notch | T2 - T3 |
Half-Life Ra-226 | 1600 years |
Tray factor | Output with tray / output with openTray factors increase with increasing energy (more transmission) |
TBI Rx Dose | 1000 - 1200 cGy |
Electron Arcs | - Velocity effect decreases surface dose, increases depth, increases x-ray contamination- Extend field width 15' to retain penumbra - Shielding thickness E/2 + 1mm Pb |
Skin TD 5/5 (100cm square) | 5000 cGy |
Controlled Area | Exposure of persons is monitored by radiation protection of 0.1 rem/week. |
Tenth Value Layer | Amount of material required to reduce beam intensity to one-tenth its original value.TVL = 3.3 HVL |
Power Equation | - Power = Voltage (current)- Measured in watts |
Median Target Dose | Value between min & max absorbed dose values within the target. |
Co-60 source size | 1.5 - 2 cm |
Exposure | Total charge of ions of one sign, liberated by photons per unit mass in air. |
Non-controlled area | Unsupervised by radiation protection = 0.01 rem/wk |
CT # Air | -1000 |
Time to decay to 10% radioactivity | 3.32 Half Lives |
Mass attenuation due to compton interactions depends on... | Electron density |
B+ decay | ... |
Phantom Scatter (Sp) | - Not directly measurable- Increases with field size |
Penumbra transmission through collimator & block | - Collimator = 0.5%- Block = 3% |
Advantages of multiple field treatments | - Maximization of target dose- Minimization of normal tissue dose - Applicability in all clinical situations |
Spatial Resolution of PET | 3-4 mm in axial plane |
CT Accuracy | 1 mm |
> 20 cm separation - what energies are used? | 10 MV or higher x-ray |
Increase filtration | - Decreased dose rate- Decreased # of lower energy photons - Increased HVL |
CT # of Water | 0 |
Co-60 Timer Error | 0.02 minutes |
Fractionation results in... | - Allows for normal tissue recovery- Causes more damage to tumor than OAR |
Mean Lethal Dose | Dose to reduce surviving fraction to 10% of irradiated cells |
IORT Single Fx Dose | 1000 - 2000 cGy |
Dose @ Dmax / Midpoint Dose | - Increase thickness, increase ratio- As E decreases, increase ratio - As field size increases, decreases ratio |
Surface Dose - 6 keV | 75% |
Planck's Constant | 6.62 x 10 ^-34 J sec |
Effects of electron contamination | - Increases surface dose- Increases dose in build-up - Increases for increasing SSD - Increases with field size |
Liver TD 5/5 (whole) | 30 Gy |
When internal conversion occurs... | - Z and A remain the same- Energy transferred to an inner shell electron, which is ejected |
Coherent Scatter | - Independent- Happens at 30 keV - 30 MeV |
Half-Life of Pd-103 | 17.0 days |
Which is hotter SSD or SAD treatments? | SAD treatments |
Contact Therapy | - 40-50 kVp- HVL < 1mm AL Dmax - surface |
Use Factor (U) | Fraction of operating time beam is directed toward the barrier. |
Coefficient of Equivalent Thickness of Lung | 0.25 cm water = 1 cm of lung |
Level of cricoid cartilage | - C6- Junction of larynx to trachea & pharynx to esophagus |
Binding energy per nucleon | Determines the stability of the nucleus. |
Half-Life of Free Neutron | 12 min |
Surface Dose - 9 MeV | 80% |
Scatter dose at any point in a phantom can be determined by... | - Scatter air ratio- Scatter maximum ratio |
Photoelectric interacting photon energy | - Equal to or just greater than binding energy of electron. |
LINAC focal spot size | 0.5 - 3.0 mm |
High Energy Beam Calibration Chamber | Farmer Type |
RBE | RELATIVE BIOLOGICAL EFFECTIVENESS- Increases with LET - Highest for charged particles |
Oxygen Enhancement Ratio | - Oxygen enhances cell kill- Greater for X-rays than neutrons - Dose without oxygen / Dose with oxygen ... to cause the same effect |
Electron Binding Energy | - Decreases with increasing shell #- Increases with increasing atomic # - Equal for the same shell for all isotopes of the same element |
Degrees C = ? Degrees F | C = 5/9 (F - 32) |
Penumbra for Co-60 | 15 mm |
SRS penumbra | - Best with Gamma Knife- Worsens with additional shots and arcs |
HVL (mm Lead) Ir-192 | 2.5 mm |
BSF Co-60 | 1.036 |
Penumbra (Cerrobend vs. MLC) | Cerrobend provides a tighter penumbra than MLC |
For dose points that lie beyond an area of inhomogeneity, the prominent effect is... | Attenuation of primary beam |
Mean energy of an electron | 2.33 MeV/cm x R50 |
CT # fat | -100 |
Paris System | - Uniform loading- Non-uniform dose - No crossing - Parallel planes |
SI unite of dose equivalent | Sievert |
Integral Dose | - Measure of total energy absorbed in the treated volume- Mass x dose (kg-rad) - Keep to minimum - Decreases as energy increases |
Xiphoid Process | T9-T10 |
LD 50/30 Humans | 450 cGy |
Monthly Linac Output | 2% |
Sievert Integral | - Calibrates brachy source doses of Cesium and RadiumINCLUDES: - Air - Inverse Square - Filtration EXCLUDES: - Self absorption - Tissue attenuation - Tissue scatter |
Colon TD 5/5 (whole | 45 Gy |
Fricke Dosimeter | - Works via calorimetry- Provides absolute dose measurement - Has a large range |
Therapeutic Ratio | - Ratio of tissue tolerance to tumor lethal dose- Less than 1 for resistant tumors - Larger than 1 for sensitive tumors - Can be modified with radiosensitizers and radioprotectors |
Paterson-Parker | - Non-uniform distribution produces uniform dose +/- 10%- Area reduced by 10% for each uncrossed end |
Surface Dose - 12 MeV electrons | 85% |
Standquist Curves | - Radiation Biology model based on skin tolerances |
I-125 dose-rate | 5-10 cGy/hr |
Monthly Flatness | 3% of 80% of field at a depth of 10 cm |
Monthly Symmetry | 2% of 80% of field at a depth of 10 cm |
Dose beyond beam edge is influenced by... | - Photon side scatter- Electron Transport - Collimator Transmission |
HDR Unit Source | 10 Ci Ir-192 |
Relationship between speed, wavelength, and frequency... | c = λν |
Curie (Ci) | 1 g Ra-226 = 1 Ci1 Ci = 3.7 x 10^10 disintegrations per second 1 mg Ra-226 = 1 mCi = 3.7 x 10^7 dps 1 Ci = 3.7 x 10^10 Bq |
Quimby | - Uniform source distribution- Along & Away - Area reduced by 7.5% for each uncrossed end |
Picket Fence Test | Checks leaf positions that are out of calibration |
Chamber measurements must be corrected for | - Saturation- Polarity - Temperature/Pressure |
Factors needed to select correct PDD for a superficial treatment | 1. HVL2. SSD |
Surface dose for a 6 MV beam | 15% - 40% depending on field size |
Ratio of scatter @ 1m to entrance dose to patient | 1:1000 |
Photon energy of Co-60 | 1.17 MeV1.33 MeV Average Energy is 1.25 MeV |
Penumbra | Distance between the 80% and the 20% IDL. |
Prostate seed RX for Pd-103 | Primary - 115 GyBoost - 90 Gy |
Compton initial interacting photon energy | Much larger then electron binding energy |
Dmax for 18 MV x-rays | 3.5 cm |
Source Inventory Contains | - Source identity- Source configuration - Source strength |
Transient Equilibrium | - Half-life of parent is not much different than half-life of daughter; occurs at about 1.5 days |
MDR Dose Rate | 200 - 1200 rad/hr |
Isotopes | - Have same # protons- Have same # electrons - Chemically identical |
HDR Dose Rate | > 1200 rad/hr |
Shape of H & D Curve | ... |
Planning Target Volume (PTV) | Includes CTV with IM + setup margin |
Output Factor -- Circle vs. Rectangle of Same Area | Output is greater for circle |
Maximum Scatter Energy | 0.511 MeV |
Dmax 1.25 MeV (Co-60) | 0.5 cm |
Speed of light (c) | 3.0 x 10^8 m/s |
Wedge Angle | WA = 90 degrees - (hinge angle/2) |
Penumbra Equation | ... |
Effective energy of an X-ray beam | - Energy of a monoenergetic beam that would give the same HVL |
Dmax of 9 MeV electron | 2.0 cm |
Esophagus TD 5/5 (whole) | 55 Gy |
I-125 decays by... | Electron Capture |
HVL (mm Lead) for Au-198 | 2.5 mm |
Photoelectric photon loses... | All energy |
Main component of a computer that processes all data... | CPU |
Hounsfield Units | ... |
Half Life Rn-222 | 3.83 days |
1mCi = ? Bq | 3.7 x 10^7 dps (Bq) |
Becquerel | 1 Bq = 1 dps = 1 atom/sec1 Ci = 3.7 x 10^10 Bq |
B- Decay | ... |
Secular Equilibrium | Half-life of parent is much greater than half-life of daughter -- occurs at about 20-25 days. |
Co-60 unit dose rate | 240 cGy/min |
Modal Target Dose | Absorbed dose that occurs most frequently within the target |
Internal Conversion | ... |
Hot spot Co-60 | 30% |
Atomic Mass # | #n + #p = # nucleons |
Point A | - 2 cm up from os, 2 cm over from tandem- Where uterine vessels cross the ureter - Prescription Point |
Half life Co-60 | 5.26 years |
Dose equivalent is measured in... | Rem % Sv |
HDR Source Positioning Accuracy | 1 mm |
Dmax for 250 keV | Surface |
SRS Dose | - Volume dependent in that bigger volumes get smaller doses- Gamma knife prescribe to 50% line - LINAC - prescribe to 80% line |
Fraction of Activity Remaining is Equal to... | ... |
HVL (mm lead) Rn-222 | 12 mm |
Radiation Exposure -- Public One-Time Max | - 0.5 rem/yr- 5 mSv/yr |
Linear Quadratic Formula | Rad Bio model based on biological effective dose (BED) |
Shoulder of cell survival curve represents... | - Repair of sublethal damage- No damage due to small radiation doses - Absent for high LET radiation |
LD 50/60 Humans | 300 rad |
Dose decrease beyond bone (Co-60) | 4% per cm |
TLDs | - Dose-rate independent- Large range - Personnel monitoring - Patient dosimetry - Must be stored properly |
Thyroid Cartilage | C4-C5 |
Ion Chamber Uses | - Surveys & Calibrations- Brachy patient survey - Source calibration - LINAC calibration - Shielding surveys |
Dose perturbation behind inhomogeneity depends on... | - Size- Shape - Electron Density - Effective Atomic # |
Cell Death | - Cell losing some specific function in case of non-proliferating cells- Caused by lethal damage to DNA & other critical structures |
TBI Dose-Rate | 5-10 cGy/min |
GI Syndrome | 10-100Gy |
Electron Energy Loss | 2 MeV/cm |
Increase in dose to tissues beyond health lungs (Co-60 γ rays) | + 4% per cm in lung |
Scintillation Detectors | - Light is proportional to E- NaI used for high count rates - NaI used for contamination surveys & thryoid uptakes - Liquid scintillation counters used for B surverys |
Transpyloric Plane | Pancreas = L1 |
Electron depth dose measurements in water are carried out by... | - Ionization chamber- Silicon diode |
X-Ray contamination of electron beams | - 2 - 5% for a 16 MeV beam- Caused by interactions with high Z components inside accelerator |
CNS Syndrome | > 50 Gy |
Mandible TD 5/5 (whole) | 60 Gy |
Physical Penumbra depends on... | - Geometric penumbra- Collimator Transmission - Lateral photon scatter (patient) - Lateral electron transport (patient) |
Dose Buildup | - Is NOT linear- Very steep in initial portion of build-up - Leads to skin sparing |
1R = ? rad | 1 R = 0.873 rad |
Increase in dose to tissue beyond healthy lung with 6MV xrays | + 2.5% per cm lung |
Brainstem TD 5/5 (whole) | 50 Gy |
Dose decrease beyond bone with 4MV photons | 3% per cm |
TD 5/5 bladder (whole) | 65 Gy |
Bits per pixel required in a CT image to represent the full range of CT numbers | 12 bits/pixel |
Dose decrease beyond bone 10MV | 2% per cm |
Photon Energy I-125 | 0.028 MeV = 28 keV |
What is the acceptable block transmission of the primary beam? | 5% |
Quimby System | - Uniform loading- Non-uniform dose - Crossing needles - Along & Away - Higher dose in center than periphery |
Water phantom for measuring isodose distribution must extend at least ___ laterally & deep beyond the field of measurement | 10 cm |
Depth Dose | % DD = Dose @ Depth / Dose @ Dmax |
Pair Production | - Depends on Z^2- Must be energy > 1.022 Mev |
Attenuation of Cerrobend is... | About 15% less than lead |
1 Gy is equal to... | - 1 J/kg- 100 rads - 1000 mGy - 1 Sv/QF |
Energies greater than 10MV PDD @ d = 10 cm decreases by ... | 2% per cm |
Hematopoietic Syndrome | - 1-10 Gy- Survivable |
HVL (mm lead) Co-60 | 11 mm |
What improves dose homogeneity for TBI? | - Using highest beam energy & beam-spoiler- AP/PA beams - Tissue compensators for extremities - Using large SSD/SAD - Adding compenating material over the lungs |
1 SV = ? rem | 1 SV = 100 rem |
T2 MRI | - No contrast- Shows swelling and edema - Swelling, edema, and CSF hows up white |
Mag Factor | Image ÷ Object |
Half Life of Au-198 | 2.7 days |
Isotone | - Same # neutrons- Different # p - Different A - Different atom |
Law of Bernie & Tribundeau | Radiosensitivity varies with:- Cell proliferation - rapidly dividing are sensitive - Future cell division - long-diving futures more sensitive - Inversely with differentiation - less differentiated are more sensitive |
Dmax with 6 MV X-rays | 1.5 cm |
Optic Nerve TD 5/5 (whole) | 50 Gy |
Heavy Charged Particles | - Maintain direction- Have a defined range - Have a Bragg curve |
Parotid Gland TD 5/5 (whole) | 32 Gy |
4 R's of Radiotherapy | 1. Repair2. Repopulation 3. Reoxygenation 4. Redistribution |
Depth of Dmax for a photon beam is approximately equal to... | - Depth at which dose & kerma are equal- Maximum range of secondary electrons - Depth at which electronic equilibrium occurs |
Avagadro's Number | 6.023 x 10^23 atoms per gram |
Superficial X-rays | - 50 to 150 kVp- HVL = 1-8 mm Al - Dmax = Surface |
Exposure Rate Constant for Au-198 | 2.3 Rcm^2 per mCi-h |
Craniospinal Couch Angle | ... |
Penumbra - Accelerator | 8 mm |
Pt B Dose | 1/3 of Pt A dose |
Shades of gray in 1 byte | 256 shades of gray |
Dmax 10 MV X-rays | 2.5 cm |
A H&D curve represents | Dependency of optical density on dose |
Facts about MRIs | - MRI signal differentiates between differences in hydrogen content.- Can image metabolic activity - Only ferromagnetic materials cause problems - Radiographs have better resolution |
Isotope | - Same # protons- Different # neutrons - Different A - Same Z - Same atom |
Pacemaker Limit | 2 Gy |
Compensators | Provide same distribution @ 1 depth- overcompensate at shallow depths - Under compensate at deep depths Preserve skin sparing 15 cm away Compensator thickness < tissue replacement thickness |
Dose Equivalent | Rem < - > Sievert |
Jugulodigastric Lymph Nodes | Superior to angle of the mandible; Drains tonsils & tongue |
PDD | Dose depth ÷ Dmax @ SSD |
Cumulative Dose in Brachytherapy | 1.44 x T 1/2 x Initial Dose rate |
Arc Therapy | - Good for small deep-seated tumors- Should not be used if: large volume, non-cylindrical, off-center - Past pointing for partial arcs - Smaller skin dose |
Effects of field blocking | - Reduces scatter volume- Reduces scatter dose @ depth - Primary dose is relatively unaffected |
Vascular Brachytherapy Dose | 15 - 20 Gy |
Energy of a Bremstrahlung X-ray | Any energy up to that of the incident electron |
Stomach TD 5/5 (whole) | 50 Gy |
DRR Quality depends on... | - Slice #- Slice thickness - Technique used - Reconstruction algorithm |
Lung TD 5/5 (whole) | 17.5 Gy |
Transmission through MLC | 2% |
Spinal cord TD 5/5 (20 cm) | 47 Gy |
Energies of Positrons & Betas | Spectrum |
Electron Range | E/2 |
Coherent Scattering | - Energy unchanged- Direction changed - Conservation of energy |
Max Radiation Exposure to Extremities for XRT workers | 50 rem/yr = 500 mSv/yr |
Point B | 2 cm up from os, 5 cm lateral to midline; represents obturator nodes. |
IORT | - 1 fx of 10 -20 Gy- 6 MeV - 21 MeV electrons - Increasing energy, flatness worsens, increased penumbra, increased x-ray contamination |
Total Skin | - 6 field Stanford technique- Shield eyes & nails - Boost perineum, shoulders, folds, breast |
CT # Lung | -400 |
How many HVLs yields less than 5% primary beam transmission? | Between 4.5 & 5 HVL |
Photon energy for Au-198 | 0.412 MeV = 412 keV |
Collimator Scatter (Sc) | - Measured in air- Increases with field size |
NaI well chamber | - Measures low level gamma rays- Used in wipe testing |
Alpha Decay | - Alpha particle = Helium nucleus = He- Usually occurs only for Z > 82 - Typical E (alpha) ~ 4-7 MeV - Each radioactive element has a discrete alpha energy |
Lowly Sensitive Cells | - Nerve- Muscle |
T1 MRI | - Tumor- With Constrast - CSF (ventricles) are black - Sagittal sinus is white due to contrast - No edema |
PET scanners detect... | Annihilation photons |
1 Gy = ? rad | 100 rad |
Co-60 decays... | - 1% per month- 13 % per year |
TDF Model | Radiation Biology model based on single exposure skin tolerances with time, dose, and fractionation factors built in. |
Distance Between 2 Points | ... |
Mean Target Dose | Mean of absorbed dose values within target |
Photon Energy (Cs - 137) | - 0.662 MeV- 662 keV |
Patterson-Parker | - Uniform Dose- Non-uniform distribution - Crossing Needles - Peripheral Loading |
Advantages of a dynamic wedge over a physical wedge | - Same depth dose as open beam- Field height not limited - No heavy lifting - Less dose outside field |
Half Value Layer | - Amount of material required to reduce beam intensity to half its original value- Monoenergetic HVL1 = HVL2 - Polyenergetic HVL1=HVL2 - HVL = .693/u |
Increase in dose to tissues beyond healthy lung - 10 MV xrays? | + 2cm per cm lung |
Isobar | - Same A- Different p - Different n - Different Atoms |
Focal spot of clinical LINAC | 3 mm |
Craniospinal Collimator Angle | ... |
B+ decay occurs in accordance with the laws of | - Electric charge conservation- Energy conservation - Momentum conservation |
Dmax for 12 MeV electron | 2.5 cm |
Exposure Rate Constant - Ir-192 | 4.69 Rcm^2 / mCi-h |
Hot Spot 10MV | 10% |
Scatter of MV beams increases with... | - Depth- Field Size - Dose Rate |
Causes of the OAR | - Flattening Filter- Photon Scatter |
Diodes | - GeLi used to identify isotopes- Semiconductors are very efficient |
Lasers & ODI | +/- 2mm |
Photon Energy for Pd-103 | 0.021 MeV = 21 keV |
Half-Value Layer/Linear Attenuation Coefficient Equation | u = .693 ÷ HVL |
Dynamic Wedge | - Moving jaw creates wedged effect- Larger wedged field sizes - Any wedge angle |
Heart TD 5/5 (whole) | 40 Gy |
Transmission through jaw | 1% |
Thoreus Filter | - Target * Tin (Sn) - removes low energy photons which would only increase skin dose (increase average effective energy) * Cu -- removes char x-rays produced by target * Al -- removes char x-rays produced by target - Patient Filter decreases output but increased effective energy, leading to better penetration |
Co-60 | - 1.25 MV- HVL 1.2 cm Pb - Dmax 0.5 cm |
Divergent blocks reduce | Transmission penumbra |
FLAIR MRI | - Swelling and edema show up white- CSF is black - Needed to differentiated between edema & CSF |
Occupancy Factor (T) | Fraction of operating time during which the area of interest is occupied by the individual |
Electron depth (80%) | E ÷ 3 |
Scattering Foil | - Primary - spreads out electrons- Secondary - flattens electron beam |
First Time Here?
Welcome to Quizlet, a fun, free place to study. Try these flashcards, find others to study, or make your own.